Screening the orbits for metal before MRI: X‑ray and ultra-low-dose CT doses compared

A phantom study published online in Radiography on 19 November 2026 measured the radiation dose to the eye from two ways of checking for a metallic foreign body in the orbit before MRI: digital radiography (X-ray) and ultra-low-dose photon-counting CT. In this model, the CT protocol gave a much lower dose to the lens and globe. It is a dose study on a phantom, not a test of diagnostic accuracy in patients.

What's new

A ferromagnetic fragment in the eye can move or heat in an MRI scanner and cause serious eye injury. When a patient's history cannot rule this out, imaging of the orbits is done first. Plain X-ray is commonly used because its dose is low, but it shows little anatomical detail. Photon-counting CT is a newer CT technology with high spatial resolution.

The team imaged a standard anthropomorphic phantom (Alderson RANDO) with both methods and measured dose with optically stimulated luminescence dosimeters at the lens and globe. Mean absorbed dose to the lens was 6.215 mGy with X-ray and 0.092 mGy with the CT protocol (a 98.5% reduction); to the globe it was 4.855 mGy and 0.087 mGy (98.2%). Estimated effective dose was 0.023 mSv for X-ray and 0.002 mSv for CT (91.3%). The tested metal fragment was visible on CT.

The authors stress that diagnostic performance has not yet been tested and that further validation is needed before routine clinical use. Results depend on the specific X-ray views, exposure settings and CT protocol used at one German centre, so they may not match other departments. Our summary is based on the published abstract, which does not list the X-ray projections used.

What it means for learners

  • Know why the orbit film is done. Pre-MRI orbit imaging is a safety check for metal, not a test for fracture. It is done when the patient's history cannot reliably rule out metal in the eye.
  • "Low dose" depends on what you compare. Plain films are usually the low-dose option, but here a modern CT protocol gave less dose to the eye. Dose depends on technique, not only on the modality. Revisit the principles in radiation dose and protection.
  • Organ dose and effective dose tell different stories. Here the reduction was 98.5% for the lens but 91.3% for effective dose. Look at which dose a study reports before comparing figures.
  • Metal shows; much else does not. For how different materials appear on X-ray, see foreign bodies on X-ray. Local MRI safety policy decides which test is used.

Source

  • Vogt ML, Centurioni G, Günthner-Lengsfeld T, Salazar Hammann A, Toussaint A, Wittig-Sauerwein A, Pham M, Weick S, Guggenberger KV. Comparison of ocular radiation doses between ultra-low-dose photon-counting computed tomography and digital radiography for orbital foreign body screening: a phantom study. Radiography. 2026;32(8):103645. Published online 19 November 2026. doi:10.1016/j.radi.2026.103645 (phantom dosimetry study; summary based on the published abstract).