IABP and pulmonary artery (Swan-Ganz) catheter position on chest X‑ray
Two devices appear on cardiac intensive care and post-cardiac-surgery chest films and are easy to overlook: the intra-aortic balloon pump (IABP), which lies in the descending thoracic aorta, and the pulmonary artery catheter (PAC, or Swan-Ganz catheter), which passes through the right heart into a pulmonary artery. Each is judged mainly by one small point on the film: the radiopaque marker at the top of the IABP and the tip of the PA catheter. If either sits too high or too low, too far in or not far enough, it can do serious harm: stroke or abdominal organ ischaemia with an IABP, lung infarction or pulmonary artery injury with a PA catheter. This page covers what each device does, the target positions and where sources differ, the common malpositions and a checklist for the film.
Patient safety
Tube and line confirmation must follow your organisation's policy. That policy usually sets who may confirm a position, which documentation is needed, and when a radiologist must review the film. This page explains the radiographic criteria. It does not authorise anyone to confirm positions.
1. The IABP: what it is and how it looks
An intra-aortic balloon pump is a catheter carrying a long balloon that sits in the descending thoracic aorta. The balloon is timed to inflate in diastole, just after the aortic valve closes, and deflate in systole; this is called counterpulsation (Rossini et al., 2021). It is used to improve coronary blood flow in patients with myocardial ischaemia and to support the circulation in selected patients with cardiogenic shock (Sakthivel et al., 2020; Rossini et al., 2021).
The usual route is a femoral artery, with the catheter advanced upwards into the descending aorta (Rossini et al., 2021). Some devices are placed from the arm, through a subclavian or axillary artery, and then the catheter comes down over the arch (Sakthivel et al., 2020; Rhodes et al., 2022). Sakthivel and colleagues stress that the reader must know which route was used, because it changes what a good position looks like and which way a malpositioned device needs to move.
Most of the catheter is radiolucent, so the device is identified by its radiopaque markers. One review describes markers at both ends of the balloon (Gambato et al., 2023), but the number and design of markers differ between models and manufacturers (Rhodes et al., 2022), so do not assume every IABP looks the same. In a femoral insertion, the marker you judge position by is the one at the upper (cranial) end, often called the tip marker.
2. IABP target position
The aim is for the top of the balloon to sit in the proximal descending thoracic aorta, just beyond the origin of the left subclavian artery, so it does not obstruct the arch vessels, while the lower end stays above the renal and splanchnic (gut) arteries (Rossini et al., 2021; Gambato et al., 2023). The left subclavian origin is not visible on a plain film, so the frontal chest X-ray uses surrogate landmarks: the top of the aortic arch (the aortic knob) and the carina.
Key measurements
- Tip about 2 cm below the left subclavian artery origin and the top of the transverse arch (Sakthivel et al., 2020), versus about 2–3 cm below the left subclavian origin (Rossini et al., 2021). [VARIES: about 2–3 cm]
- On the frontal film, one review places a correctly positioned IABP at about the level of the aortopulmonary window, just under the arch (Gambato et al., 2023).
- Using the carina: in a CT-based study, a tip about 2 cm above the carina would have been 1.5–3.5 cm below the left subclavian origin in 95.3% of patients (Kim et al., 2007).
Pearl: why the carina helps
The distance from the aortic knob to the left subclavian artery varies. In Kim and colleagues' CT study, a tip placed at the aortic knob would have been too close to the left subclavian origin in 16% of patients. The authors suggested that the carina may be a more reliable landmark. Checking the marker against both the knob and the carina is a sensible habit.
3. IABP malposition and complications
| Problem | What you see | Why it matters |
|---|---|---|
| Too high (too proximal) | Tip marker at or above the top of the aortic knob, or in the line of a great vessel | A tip placed too proximally can impede cerebral blood flow and cause stroke (Sakthivel et al., 2020). |
| Too low (too distal) | Tip marker well below the arch, for example over the mid descending aorta; one published example had the marker at the sixth intercostal space before repositioning to the fourth (Gambato et al., 2023) | Less effective support for the heart and a risk of renal ischaemia (Sakthivel et al., 2020); the lower balloon may obstruct the splanchnic vessels (Gambato et al., 2023). |
| Arm-route device not advanced enough | With a subclavian or axillary insertion, the marker over a great vessel above the arch; Sakthivel and colleagues show one at the expected position of the right innominate artery | Needs advancing into the descending aorta. The lower end of an arm-route device may be off the bottom of a chest film (Sakthivel et al., 2020). |
Pitfall: anatomy that fools the landmarks
A marker in an unexpected place is not always a misplaced device. In a published case, an IABP inserted from the femoral artery projected over the right hilum, raising concern that it had entered a vein or the pulmonary artery. Earlier imaging showed the patient had a right-sided aortic arch (Mahavar et al., 2024). Look at the arch side and the course of the descending aorta, compare with previous imaging, and seek senior or radiology advice rather than assuming.
Aortic dissection is a recognised serious complication of IABPs. It is a clinical emergency diagnosed with CT angiography, not on the chest film; the plain film's main job is to show malposition (Gambato et al., 2023).
4. The PA catheter: what it is and how it looks
A pulmonary artery catheter is a long, balloon-tipped catheter, usually inserted through the right internal jugular vein (Gambato et al., 2023), though other veins, including the femoral vein, are also used (Sakthivel et al., 2020). It passes through the right atrium, right ventricle and pulmonary valve into a pulmonary artery. It is mainly used to measure pulmonary artery pressures, and is sometimes used to deliver drugs such as thrombolytic infusions (Sakthivel et al., 2020).
On a frontal film the catheter is a continuous radiopaque line. Follow it from the neck down the line of the superior vena cava, through the right heart in a curve, up through the pulmonary outflow tract and out into the right or left pulmonary artery. It usually goes right: in a study of 135 cardiac surgery patients checked with transoesophageal echocardiography, the catheter was in the right pulmonary artery in 96% and the left in 3% (Tempe et al., 2015). After cardiac surgery a PA catheter often lies alongside a central venous line and pacing wires, so trace each line separately.
5. PA catheter target zone
Key measurements
- The tip should end centrally in the right or left main pulmonary artery, ideally proximal to the interlobar pulmonary artery (Sakthivel et al., 2020).
- The catheter should never cross the hilum (Sakthivel et al., 2020).
- A tip that has not yet reached a main pulmonary artery, for example one still in the right ventricle, is short of this target zone.
Why central? A catheter that drifts into a small peripheral branch can block it, and peripheral wedging of the tip, balloon over-inflation and sudden distal migration are among the proposed mechanisms of pulmonary artery injury. A case report and review lists keeping the catheter near the proximal pulmonary artery and limiting balloon inflation among the preventive measures (Nakamura et al., 2025). The frontal film is a two-dimensional estimate, and local policy decides what is acceptable.
6. PA catheter malposition and complications
- Tip too distal. Sakthivel and colleagues illustrate catheters ending in right lower lobe segmental branches after both internal jugular and femoral insertion. A distal tip risks pulmonary infarction or haemorrhage.
- Infarction or haemorrhage beyond the tip. Look for new wedge-shaped peripheral consolidation distal to the catheter tip (Sakthivel et al., 2020).
- Pulmonary artery injury. Rare but often fatal: one review quotes an incidence of 0.1–0.8% and mortality of 45–60%, with over 90% of injuries arising in the right lung (Nakamura et al., 2025). Perforation of the right atrium or pulmonary artery is diagnosed mainly on clinical features and echocardiography rather than on the chest film (Gambato et al., 2023).
- Looping and knotting. Spare catheter can loop in the right heart, risking valve damage or arrhythmia (Sakthivel et al., 2020); the chest film is useful for showing loops and knots (Gambato et al., 2023).
- Insertion complications. As with any central venous access, look for a pneumothorax and any new pleural or mediastinal collection (see mediastinum and hila).
Pitfall: the film shows projection, not function
A frontal film shows where the tip projects, not whether the balloon is deflated or whether the readings are valid. Report what you see and let the team interpret it together with the pressure traces and local policy.
7. A systematic review checklist
- Identify every device. Endotracheal tube (see ETT position), central lines, PA catheter, IABP markers, pacing wires, drains, sternal wires. Look specifically for an IABP marker near the arch: it is small and easy to miss.
- IABP: establish the route. Femoral (comes up from below) or subclavian/axillary (comes down over the arch)? Check the request or the notes.
- IABP: locate the tip marker against the top of the aortic knob and the carina. Is it just below the arch and above the carina?
- IABP: check the aorta. Which side is the arch? Does the device follow the line of the descending aorta?
- PA catheter: trace the whole course from the neck to the tip. Any loop in the right atrium or ventricle, or a knot?
- PA catheter: locate the tip. In a main pulmonary artery, inside the hilum?
- Look beyond the tip. New wedge-shaped peripheral opacity? New pleural fluid, pneumothorax or mediastinal widening?
- Compare with the previous film. A change in position between films matters more than any single number.
- Document and escalate according to local policy.
Red flags: act now
Escalate immediately
- An IABP marker at or above the top of the aortic knob, or in the line of a great vessel.
- An IABP marker well down the descending aorta (risk of renal and gut ischaemia and poor support).
- A device whose course does not follow the expected vessel, unless a known anatomical variant explains it.
- A PA catheter tip beyond the hilum, or a new wedge-shaped opacity distal to it.
- New haemoptysis or blood in the endotracheal tube in a patient with a PA catheter: possible pulmonary artery injury.
- A looped or knotted PA catheter, or a new pneumothorax after insertion.
What to focus on, by role
| Role | Priorities |
|---|---|
| Medical student | Recognise both devices on a cardiac ICU film. Know that the IABP marker belongs just below the aortic arch and above the carina, and that a PA catheter tip belongs in a main pulmonary artery inside the hilum. |
| Nurse | Know that both devices can migrate. Report any change in the patient after insertion or repositioning, and check that the position has been confirmed as local policy requires. |
| Junior doctor | Document the IABP marker against the aortic knob and carina, the insertion route, and the PA tip relative to the hilum. Compare with the previous film and escalate any change. |
| Radiographer | Include the arch, carina and both hila, and minimise rotation, which distorts the landmarks. Follow local protocol on extra coverage for an IABP. |
| Advanced / FRCR | Arm-route IABP appearances, marker differences between models, the evidence for the carina versus the aortic knob, and anatomical variants such as a right-sided arch. |
See also: central line and PICC position, endotracheal tube position, pacemakers and cardiac devices, mediastinum and hila and the glossary.
Practise on lines and tubes films →
References
- Sakthivel MK, Bosemani T, Bacchus L, Pamuklar E. Malpositioned lines and tubes on chest radiograph: a concise pictorial review. J Clin Imaging Sci 2020;10:66. doi:10.25259/JCIS_170_2020 (open access)
- Kim JT, Lee JR, Kim JK, et al. The carina as a useful radiographic landmark for positioning the intraaortic balloon pump. Anesth Analg 2007;105(3):735–738. doi:10.1213/01.ane.0000278086.23266.35
- Rossini R, Valente S, Colivicchi F, et al. ANMCO position paper: role of intra-aortic balloon pump in patients with acute advanced heart failure and cardiogenic shock. Eur Heart J Suppl 2021;23(Suppl C):C204–C220. doi:10.1093/eurheartj/suab074 (open access)
- Gambato M, Scotti N, Borsari G, et al. Chest X-ray interpretation: detecting devices and device-related complications. Diagnostics (Basel) 2023;13(4):599. doi:10.3390/diagnostics13040599 (open access)
- Rhodes NG, Johnson TF, Boyum JH, et al. Radiology of intra-aortic balloon pump catheters. Radiol Cardiothorac Imaging 2022;4(2):e210120. doi:10.1148/ryct.210120
- Mahavar R, Singh A, Mehta Y. Assessing the position of the intra-aortic balloon pump catheter tip in right-sided aortic arch. Ann Card Anaesth 2024;27(3):290–292. doi:10.4103/aca.aca_6_24
- Tempe DK, Batra UB, Datt V, et al. Where does the pulmonary artery catheter float: transesophageal echocardiography evaluation. Ann Card Anaesth 2015;18(4):491–494. doi:10.4103/0971-9784.166450
- Nakamura Y, Saku K, Homma S, et al. Pulmonary artery rupture by pulmonary artery catheter in cardiac surgery: a case report and review of literature. Front Cardiovasc Med 2025;12:1567723. doi:10.3389/fcvm.2025.1567723 (open access)
For education only. Not a substitute for a formal report, senior review or your local policy. Never use this page to make decisions about a real patient.